54
3
Making a Measurement
Staring a Measurement Cycle (triggering the instrument)
Staring a Measurement Cycle (triggering the instrument)
Configuring the Instrument to Automatically Perform Continuous
Measurement
Step 1.
Select a measurement to be triggered using the
Step 2.
Set the trigger source of the triggered measurement to Internal using the
:TRIGger:xx:SOURce command.
Step 3.
Set the continuous initiation mode of the enabled triggered measurement using the
:INITiate:xx:CONTinuous command.
Starting Measurement on Demand
Step 1.
Select a measurement to be triggered using the
Step 2.
Set the trigger source of the triggered measurement to Bus using the :TRIGger:xx:SOURce
command.
Step 3.
Set the continuous initiation mode of the enabled triggered measurement using the
:INITiate:xx:CONTinuous command.
Step 4.
Trigger the instrument when you want to perform measurement. An external controller can
trigger the instrument by only a single command.
Step 5.
To start the next measurement cycle, repeat step 4.
Command
Applicable
trigger source
Bus trigger only
Содержание E5052A
Страница 4: ...4 ...
Страница 28: ...28 1 Making Effective Use of This Manual How to Use This Manual ...
Страница 51: ...3 51 Making a Measurement Trigger System 3 Making a Measurement Figure 3 1 Trigger system ...
Страница 58: ...58 3 Making a Measurement Waiting for End of measurement ...
Страница 76: ...76 4 Reading Writing Measurement Data Entering Data in a Trace ...
Страница 82: ...82 Chapter5 Saving and Recalling File Management Managing Files ...
Страница 476: ...476 AppendixB Status Reporting System Status Register Structure Figure B 3 Status register structure 1 of 6 ...
Страница 478: ...478 AppendixB Status Reporting System Status Register Structure Figure B 5 Status register structure 3 of 6 ...
Страница 480: ...480 AppendixB Status Reporting System Status Register Structure Figure B 7 Status register structure 5 of 6 ...