488
Appendix B
Status Reporting System
Status Register Structure
register.
Issuing the
command will clear all of the bits in the questionable limit PN1 status
event register.
Issuing the
command will clear all of the bits in the questionable limit SP1 status
event register.
Issuing the
command will clear all of the bits in the questionable limit FP1 status
Table B-12
Status Bit Definitions of Questionable Limit PN1 Status Event Register
Bit
Position
Name
Description
0
Not used
Always 0.
1
Trace 1 Limit Test Fail
Set to “0” when a measurement cycle begins;
set to “1” when the measurement cycle finishes and
returns “fail” as the limit test result for trace 1.
2-15
Not used
Always 0.
Table B-13
Status Bit Definitions of Questionable Limit SP1 Status Event Register
Bit
Position
Name
Description
0
Not used
Always 0.
1
Trace 1 Limit Test Fail
Set to “0” when a measurement cycle begins;
set to “1” when the measurement cycle finishes and
returns “fail” as the limit test result for trace 1.
2-15
Not used
Always 0.
Table B-14
Status Bit Definitions of Questionable Limit FP1 Status Event Register
Bit
Position
Name
Description
0
Not used
Always 0.
1
Trace 1 Limit Test Fail
Set to “0” when a measurement cycle begins;
after a measurement cycle bigins, and when “fail” is
returned for trace 1, “1” will be set.
2
Trace 2 Limit Test Fail
Set to “0” when a measurement cycle begins;
after a measurement cycle bigins, and when “fail” is
returned for trace 2, “1” will be set.
3
Trace 3 Limit Test Fail
Set to “0” when a measurement cycle begins;
after a measurement cycle bigins, and when “fail” is
returned for trace 3, “1” will be set.
4
Trace 4 Limit Test Fail
Set to “0” when a measurement cycle begins;
after a measurement cycle bigins, and when “fail” is
returned for trace 4, “1” will be set.
5-15
Not used
Always 0.
Содержание E5052A
Страница 4: ...4 ...
Страница 28: ...28 1 Making Effective Use of This Manual How to Use This Manual ...
Страница 51: ...3 51 Making a Measurement Trigger System 3 Making a Measurement Figure 3 1 Trigger system ...
Страница 58: ...58 3 Making a Measurement Waiting for End of measurement ...
Страница 76: ...76 4 Reading Writing Measurement Data Entering Data in a Trace ...
Страница 82: ...82 Chapter5 Saving and Recalling File Management Managing Files ...
Страница 476: ...476 AppendixB Status Reporting System Status Register Structure Figure B 3 Status register structure 1 of 6 ...
Страница 478: ...478 AppendixB Status Reporting System Status Register Structure Figure B 5 Status register structure 3 of 6 ...
Страница 480: ...480 AppendixB Status Reporting System Status Register Structure Figure B 7 Status register structure 5 of 6 ...