2
Overview of Remote Control
LAN Remote Control System
Step 2.
In the Instrument Manager setting screen, click
Add...
.
Step 3.
The Instrument Properties setting screen appears. Make the settings as follows: Name (1 in
the figure below):
SICL_LAN
(you can specify any name), Interface (2 in figure):
GPIB
,
and Address (3 in figure):
917
(for the E5052A, fixed to 9; 17 is the address of the
SICL-LAN server). Then, click Gateway:
This host
. The Select or Enter Gateway setting
screen appears. Enter the IP address or host name of the E5052A (4 in figure).
Figure 2-3 shows an example of control using the I/O interface set by the above procedure.
Figure 2-3
Example of control using Agilent VEE