Controls
that
Affect
the
Netw
ork
Analyzer
Normal
Step.
Press
4
SYSTEM
5
NNNNNNNNNNNNNN
MORE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SYSTEM
PHASELOCK
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STEP
TYPE:
NORMAL
to
select
Normal
Step
mo
de.
In
normal-step,
the
net
w
ork
analyzer
tunes
to
a
frequency
and
measures
all
necessary
parameters
b efore
breaking
phaselo c
k
and
tuning
to
the
next
frequency
.
The
net
w
ork
analyzer
go
es
through
a
complete
phaselo c
k
sequence
at
eac
h
frequency
step.
This
metho d
of
phaselo c
k
requires
a
softw
are
handshak
e
only
(o
ccurs
through
the
System
Bus).
No
other
external
connectors
b et
w
een
the
source
and
net
w
ork
analyzer
are
required,
and
GPIB
extenders
can
b e
used.
Quick
Step
Mode.
Press
4
SYSTEM
5
NNNNNNNNNNNNNN
MORE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SYSTEM
PHASELOCK
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STEP
TYPE:
QUICK
to
select
Quic
k
Step
mo
de.
In
Quic
k
Step,
the
source
is
tuned
from
one
measuremen
t
to
the
next
in
small
steps
allo wing
the
receiv
er
to
track
the
source
without
breaking
phaselo c
k.
Since
the
time
required
for
the
receiv
er
to
ac
hiev
e
phaselo c
k
is
a
ma
jor
p ortion
of
the
measuremen
t
cycle
time
for
eac
h
p oint,
Quic
k
Step
mo
de
increases
the
sp eed
of
Step
sw
eep
measuremen
ts
up
to
six
times
with
no
p enalty
in
the
frequency
accuracy
and
rep eatabilit
y
sp ecications.
Quic
k
Step
requires
a
compatible
836xx
source.
The
A
gilent
8510C
On-Site
Servic
e
Manual
pro
vides
a
list
of
compatible
836xx
sources.
Note
Quic
k
Step
mo
de
do es
not
function
in
a
system
that
uses
m
ultiple
sources.
The
k
ey
attributes
of
the
quic
k-step
phaselo c
k
metho d
are:
Eac
h
data
acquisition
p oint
is
fully
syn
thesized.
The
source
is
\tuned"
from
p oint-to-p
oint;
it
do es
not
break
phaselo c
k.
The
net
w
ork
analyzer
remains
phaselo c
k
ed
to
the
source
except
at
the
source
bandcross
p oints
or
when
the
test
VTO
needs
to
reset.
The
net
w
ork
analyzer
and
source
require
t
w
o
BNC
connections,
describ ed
b elow.
T
ypically
(dep ends
on
a
v
eraging),
increased
data
acquisition
sp eed
(six
times
improv
emen
t)
is
ac
hiev
ed
b
y
this
metho d
of
phaselo c
k.
The
8510C
uses
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STEP
TYPE:
NORMAL
if
the
source
is
not
compatible
with
quic
k
step.
Agilent
8340/41
sources
are
NOT
compatible
with
quic
k
step.
T
o
use
Quic
k
Step
mo
de
with
an
Agilent
836xx
RF
source:
1.
Connect
the
net
w
ork
analyzer's
TRIGGER
IN
BNC
to
the
source's
TRIGGER
OUT
BNC.
2.
Connect
the
net
w
ork
analyzer's
STOP
SWP
BNC
to
the
source's
STOP
SWEEP
BNC.
3.
Press
4
SYSTEM
5
NNNNNNNNNNNNNN
MORE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SYSTEM
PHASELOCK
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STEP
TYPE:
QUICK
.
Lock
Speed
Press
4
SYSTEM
5
NNNNNNNNNNNNNN
MORE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SYSTEM
PHASELOCK
to
access
the
Lo c
k
Sp eed
con
trols.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOCK
SPEED
NORMAL
pro
vides
the
b est
frequency
accuracy
when
using
Step,
Single
P
oint,
or
F
requency
List
mo
des.
Lo c
k
Sp eed
Normal
is
the
factory
default
setting.
This
feature
has
no
eect
in
Ramp
Sw
eep
mo
de.
5-4
Using
System
Functions
Содержание 8510C
Страница 7: ...vii ...
Страница 29: ......
Страница 61: ......
Страница 73: ...Network Analyzer System Description Figure 3 7 Recommended Typical Test Setups 3 12 Principles of Operation ...
Страница 98: ...DISPLAY Functions Figure 4 9 Display Menu Showing Trace Memory Locations Menu Measurement Controls 4 19 ...
Страница 105: ...FORMAT Functions Figure 4 12 Format Function Block and Format Menu 4 26 Measurement Controls ...
Страница 106: ...FORMAT Functions Figure 4 13 Format Selections 1 of 2 Measurement Controls 4 27 ...
Страница 107: ...FORMAT Functions Figure 4 14 Format Selections 2 of 2 4 28 Measurement Controls ...
Страница 122: ...PARAMETER Functions Figure 4 25 Redefine Parameter Menu Structure Measurement Controls 4 43 ...
Страница 175: ......
Страница 203: ......
Страница 223: ...Measurement Calibration Figure 8 2 Cal Type Selections 8 8 Calibrating for System Measurements ...
Страница 267: ...Figure 9 14 Typical Group Delay and Deviation from Ideal Phase Displays 9 14 Transmission Measurements ...
Страница 273: ......
Страница 297: ......
Страница 305: ......
Страница 391: ......