Service Release for Revision A.03.81
Problem Description:
In case of using E5029H Opt.015, when the measurement interval is more than about 1
minute, next first measurement result is not good. After one bad measurement result, the
result becomes good if the interval is less than about 1 minute.
Cause: At the revision 3.81 and below, the SR1750 was set at IDLE mode at unload
position. This causes the problem. At this service release, the SR1750 was set at SLEEP
mode at unload position.