200
Chapter 3
VEE Measurement Program Operation
Measurement
Side Erase Measurement
Overview
This section describes the side erase measurement. This measurement is performed by a
spectrum analyzer or a spectrum measurement module.
Measurement
Procedure
1. To open the side erase menu, click on the [Side Erase/Reading] [Sid Erase] button from
the Measure menu.
Figure 3-121
Side Erase
NOTE
Before you execute this measurement, you should set an appropriate input range or execute
the Auto Configuration to set the input range automatically.
2. Set the parameters, click on each button to enter a new value.
Channel Bit Rate
Specifies the speed at which data bits are
written on the media per second [bps].
Write Current
Specifies the current applied to the write head.
Sense Current
Specifies the current applied to the read head.
Data Pattern
Selects the data pattern for this measurement.
Average
Specifies the number of revolutions to get the
average results.
Track #
Specifies the track number to be tested.
Center Freq.
Specifies the measurement frequency of the
spectrum analyzer. This parameter can only be
used if [Freq. Mode] is set to “Manual”.