180
Chapter 3
VEE Measurement Program Operation
Measurement
[Data Rate Sweep] BER Menu
1.
Click on [BER] button to open the BER measurement menu of the data rate sweep.
[Data Rate Sweep] > [BER] Menu
2.
Set the parameters, click on each button to enter a new value.
Meas. Bytes
Specifies the number of data bytes to be
measured.
BER Pattern #
Specifies the pattern number to be used for bit
error rate measurement. The data pattern of
the selected data pattern number must be
specified in the [Pattern Config.] menu in
advance.
Average
Specifies the number of tests to get the
average result. Average is equal to the number
of times a measurement sequence is repeated.
Refer to “hpe5022_BER_berMeas
ByteCount” function of the programming
manual for details.
Track #
Specifies the track number to be tested.
Meas. Type
Specifies the type of measurement sequence.
1)[Erase Write Meas.] means to erase, write
data then measure. 2)[Write Meas.] means to
write data, then measure. 3) [Meas.] means to
measure only.
Start Data Rate
Specifies the starting point of the data rate.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com