Agilent 4288A 1KHZ/1MHZ CAPACITANCE METER Programming Manual
Manual Change
Agilent Part No. N/A
Sep 2007
Change 1
Following command is available to use on firmware revision 1.30 or later.
:CORR:COLL:LOAD:STAN3:RANG:AUTO
Syntax:
[:SENSe]:CORRection:COLLect:LOAD:STANdard3:RANGe:AUTO {ON|OFF|1|0}
[:SENSe]:CORRection:COLLect:LOAD:STANdard3:RANGe:AUTO?
Description:
Sets the Auto Range value as ON/OFF when Load Standard measurement is performed. This command does not
influence the Load Standard measurement and only toggles the auto range setting. The Auto Range ON/OFF
setting for Load Standard measurement is saved in backup memory and can also be saved and restored using
SAVE/RECALL function
Parameter
Description
ON or 1 (Initial
setting)
Measures the load standard with auto range.
OFF or 0
Measures the load standard at measurement range which
has been set just before the load standard measurement is
performed (auto range function is invalid).
Related commands:
:CORR:COLL on page 157
:RANG on page 185
:SYST:PRES on page 194
Equivalent key sequence:
No equivalent key can be used on front panel.
Содержание 4288A
Страница 18: ...16 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 102: ...100 Chapter8 Avoiding Mistakes Related to Work and Daily Checks Daily Checks executing the self test ...
Страница 212: ...210 AppendixA Manual Changes Manual Changes ...
Страница 248: ...246 AppendixB Information for Replacing 4278A with 4288A Comparison of Interfaces ...
Страница 258: ...256 AppendixC Status Reporting System Using the Status Reporting System ...
Страница 282: ...280 AppendixG 4268A vs 4288A GPIB Command Correspondence Table 4268A vs 4288A GPIB Command Correspondence Table ...
Страница 288: ...286 Index Index ...