78
Appendix A
Agilent 4072/4073 Preinstallation Guide, Edition 4
Confirming the System Power Safety
Confirming PDU and EMO Operation
Confirming PDU and EMO Operation
When you confirm the PDU and EMO operation, use the following procedure.
1. Disconnect all power cords from the power outlets in the system cabinet, and turn off all the instruments,
controller, and peripherals.
2. Set the following:
Switch board breaker
Off position
Emergency breaker
Off position
Disconnect device
Off position
EMO button
Out position
EMO protector
In position
Fan protectors
In position
Power outlet protector
In position
To set the EMO button on the EMO panel to the
out position
, rotate the switch in a clockwise direction.
The position of the EMO, cabinet fan, and testhead fan protectors are exchanged by pressing the protector
button.
3. For the PDU’s 200 V option, check the ground-fault circuit interrupter (GFCI) as follows:
a. Turn the handle of the disconnect device to set the
on position
. When stopping the handle turning, rotate the
door lock release screw counterclockwise then set the disconnect device to on.
b. Confirm that the white push-button, labeled “LEAKAGE INDICATOR”, is in its normal position
(
in position
).
c. Press the gray push-button to force earth leakage current. Confirm that the breaker behind the disconnect
device turns off (moved to middle position), and the leakage indicator has moved to the
out position
to
indicate the earth leakage condition.
d. Set the breaker behind the disconnect device to the
on position
(reset the leakage indicator) as follows:
i.
Set the disconnect device to the
off position
. The GFCI indicator moves to the
in position
.
ii. Set the disconnect device to the
on position
.
Содержание 4072A Advanced
Страница 5: ...5 Agilent 4072 4073 Preinstallation Guide Edition 4 1 General Information ...
Страница 10: ...10 Chapter 1 Agilent 4072 4073 Preinstallation Guide Edition 4 General Information Installation Overview ...
Страница 11: ...11 Agilent 4072 4073 Preinstallation Guide Edition 4 2 Site Preparation ...
Страница 42: ...42 Chapter 2 Agilent 4072 4073 Preinstallation Guide Edition 4 Site Preparation Site Preparation Checklist ...
Страница 43: ...43 Agilent 4072 4073 Preinstallation Guide Edition 4 3 Wafer Probers and Connection with Testhead ...
Страница 57: ...57 Agilent 4072 4073 Preinstallation Guide Edition 4 4 Receiving ...
Страница 71: ...71 Agilent 4072 4073 Preinstallation Guide Edition 4 A Confirming the System Power Safety ...
Страница 83: ......