Chapter 2
27
Agilent 4072/4073 Preinstallation Guide, Edition 4
Site Preparation
Electrical Requirements
Additional power outlets
To provide power for test/maintenance equipment and other equipment not included in the system, install
additional power outlets (two power outlets, 15 A) at the site. Show the locations of all additional power outlets on
your site layout drawing and include their potential load in your calculation for total site power. Additional power
outlets must be supplied with grounding circuits intact and must be used when connecting additional equipment.