Appendix A
75
Agilent 4072/4073 Preinstallation Guide, Edition 4
Confirming the System Power Safety
Connecting Site Power Line
9. (For flexible cable,) fasten the cable clamp
securely.
10. Reinstall the disconnect device cover (removed in
step 5).
11. Reinstall the PDU side cover (removed in step 4).
12. Connect the main power cable to switchboard of
the installation site.
cable clamp
disconnect
device cover
PDU side cover