28
Chapter 2
Agilent 4072/4073 Preinstallation Guide, Edition 4
Site Preparation
Environmental Considerations
Environmental Considerations
To ensure optimum 4072/4073 tester accuracy and performance, be sure to comply with the following
environmental conditions.
Temperature and humidity
The measurement accuracy is specified at the following temperature and humidity:
Temperature:
23 °C ± 5 °C (73 °F ± 9 °F)
Humidity:
5% to 70% RH (The 4073A/4072B/4073B must be free of condensing.)
NOTE
After the SMU calibration, the temperature change must be less than 3 °C.
In addition to the above temperature and humidity conditions, Agilent Technologies recommends
that the following conditions be satisfied for measuring low-current and low-voltage with the
4073A/B:
Temperature change:
Within ± 1 °C after the calibration
Temperature change period:
≥
10 minutes
Humidity:
≤
50%
The 4072/4073 must be operated under the following conditions:
For the 4072A:
Operating temperature:
5 °C to 40 °C (41 °F to 104 °F)
Storage temperature:
−
20 °C to 60 °C (
−
4 °F to 140 °F)
For the 4073A:
Operating temperature:
5 °C to 35 °C (41 °F to 95 °F)
Storage temperature:
−
20 °C to 50 °C (
−
4 °F to 122 °F)
Storage humidity:
≤
80%
For the 4072B/4073B:
Operating temperature:
5 °C to 30 °C (41 °F to 86 °F)
Storage temperature:
−
20 °C to 50 °C (
−
4 °F to 122 °F)
Storage humidity:
≤
80%
NOTE
When using the flat panel display with the 4072/4073, the following conditions must be satisfied,
in addition to the conditions shown above.
Operating humidity:
20% to 70% RH (no-condensing)
Storage humidity range:
20% to 85% (
−
20 °C to 50 °C,
−
4 °F to 122 °F)
20% to 55% (50 °C to 60 °C, 122 °F to 140 °F)
Содержание 4072A Advanced
Страница 5: ...5 Agilent 4072 4073 Preinstallation Guide Edition 4 1 General Information ...
Страница 10: ...10 Chapter 1 Agilent 4072 4073 Preinstallation Guide Edition 4 General Information Installation Overview ...
Страница 11: ...11 Agilent 4072 4073 Preinstallation Guide Edition 4 2 Site Preparation ...
Страница 42: ...42 Chapter 2 Agilent 4072 4073 Preinstallation Guide Edition 4 Site Preparation Site Preparation Checklist ...
Страница 43: ...43 Agilent 4072 4073 Preinstallation Guide Edition 4 3 Wafer Probers and Connection with Testhead ...
Страница 57: ...57 Agilent 4072 4073 Preinstallation Guide Edition 4 4 Receiving ...
Страница 71: ...71 Agilent 4072 4073 Preinstallation Guide Edition 4 A Confirming the System Power Safety ...
Страница 83: ......