Agilent 1260 Infinity Quaternary LC VL Manual and Quick Reference
4
System Setup and Installation
Installing the Modules
Integration Into the Network
For network integration of your system refer to user manuals of your modules
(chapter
LAN Configuration
).
Table 13
Choice of Priming Solvents for Different Purposes
Activity
Solvent
Comments
After an installation
When switching between
reverse phase and normal
phase (both times)
Isopropanol
Isopropanol
Best solvent to flush air out of
the system
Miscible with almost all
solvents
After an installation
Ethanol or methanol
Alternative to isopropanol
(second choice) if no
isopropanol is available
To clean the system when
using buffers
After changing aqueous
solvents
HPLC grade water
HPLC grade water
Best solvent to re-dissolve
buffer crystals
Best solvent to re-dissolve
buffer crystals
After the installation of normal
phase seals (P/N 0905-1420)
5% isopropanol
Good wetting properties