Agilent 1260 Infinity Quaternary LC VL Manual and Quick Reference
53
Optimization of the Agilent 1260 Infinity Quaternary LC VL
3
Optimizing the Pump
3
Your pump must be connected to a data system or Instant Pilot with which
the pressure and %-ripple can be monitored, otherwhise connect a signal
cable between the pressure output of the pump and a recording device (for
example, 339X integrator) and set parameters.
Zero 50 %
Att 2^3 Chart
Speed 10 cm/min
4
Start the recording device with the plot mode.
5
Starting with a compressibility setting of 10 × 10
-6
/bar increase the value
in steps of 10. Re-zero the integrator as required. The compressibility
compensation setting that generates the smallest pressure ripple is the
optimum value for your solvent composition.
Table 9
Solvent Compressibility
Solvent (pure)
Compressibility (10-6/bar)
Acetone
126
Acetonitrile
115
Benzene
95
Carbon tetrachloride
110
Chloroform
100
Cyclohexane
118
Ethanol
114
Ethyl acetate
104
Heptane
120
Hexane
150
Isobutanol
100
Isopropanol 100
Methanol
120
1-Propanol
100
Toluene
87
Water
46