Agilent 1260 Infinity Quaternary LC VL Manual and Quick Reference
33
Introduction
1
Detector
Lamp
The light source for the UV wavelength range is a deuterium lamp. As a result
of plasma discharge in a low pressure deuterium gas, the lamp emits light over
the 190 – 600 nm wavelength range.
The lamp has an integrated RFID tag that contains the lamp specific
information (e.g. part number, burn time, ...). A RFID tag reader reads out this
information and transfers it to the user interface.
Source Lens Assembly
The source lens receives the light from the deuterium lamp and focuses it onto
the entrance slit.
Entrance Slit Assembly
The entrance slit assembly has an exchangeable slit. The standard one has a
1-mm slit. For replacement and calibration purposes to optimize the
alignment, a slit with a hole is needed.