34
Agilent 1260 Infinity Quaternary LC VL Manual and Quick Reference
1
Introduction
Detector
Filter Assembly
The filter assembly is electromechanically actuated. During wavelength
calibrations it moves into the light path.
Figure 17
Filter Assemby
The filter assembly has two filters installed and is processor-controlled.
A photo sensor determines the correct position.
Mirror Assemblies M1 and M2
The instrument contains two spherical mirrors (M1 and M2). The beam
adjustable is vertically and horizontally. Both mirrors are identical.
OPEN
nothing in light path at
λ
< 370 nm
CUTOFF
cut off filter in light path at
λ
> 370 nm
HOLMIUM
holmium oxide filter for wavelength check
SHUTTER
for measurement of dark current of photo
diodes
8jid[[;^aiZg
H]jiiZg
8jid[[;^aiZgeajh=dab^jbDm^YZ;^aiZg