3
Optimization of the Agilent 1260 Infinity Quaternary LC VL
Optimizing the Autosampler
For example:
Outside wash of needle in vial 7 before injection
Injector program:
Draw x.x (y) µl from sample
NEEDLE WASH vial 7
Inject
Wait (equilibration time - see text above)
Valve bypass
Wait 0.2 min
Valve mainpass
Valve bypass
Valve mainpass
General Recommendation to Lowest Carry-over
•
For samples where needle outside cannot be cleaned sufficiently with water
or alcohol use wash vials with an appropriate solvent. Using an injector
program and several wash vials can be used for cleaning.
In case the needle seat has got contaminated and carry-over is significantly
higher than expected, the following procedure can be used to clean the needle
seat:
•
Go to
MORE INJECTOR
and set needle to home position.
•
Pipette an appropriate solvent on to the needle seat. The solvent should be
able to dissolve the contamination. If this is not known use 2 or 3 solvents
of different polarity. Use several milliliters to clean the seat.
•
Clean the needle seat with a tissue and remove all liquid from it.
• RESET
the injector.
N O T E
Overlapped injection together with additional injection valve switching is not possible.