Agilent 1200 Series Autosampler Reference Manual
167
Repairing the Autosampler
4
SUD Board
Frequency
When defective.
Tools required
Pozidrive No. 1 screwdriver.
Parts required
SUD board G1313-66503.
Preparations for this
procedure
Remove the top cover (
Remove the transport assembly (
Remove the sampling unit (
).
C A U T I O N
Take care not to damage the flex board when removing the SUD board.
1 Disconnect all connectors from the SUD board.
2 Remove the two fixing screws, and lift out the board.
Fixing screws
Содержание 1200 series
Страница 1: ...A Agilent 1200 Series Standard and Preparative Autosamplers Reference Manual ...
Страница 12: ...12 Agilent 1200 Series Autosampler Reference Manual Contents ...
Страница 42: ...42 Agilent 1200 Series Autosampler Reference Manual 1 Installing the Autosampler ...
Страница 104: ...104 Agilent 1200 Series Autosampler Reference Manual 3 Troubleshooting and Test Functions ...
Страница 108: ...108 Agilent 1200 Series Autosampler Reference Manual 4 Repairing the Autosampler Figure 14 Using the ESD Strap ...
Страница 174: ...174 Agilent 1200 Series Autosampler Reference Manual 4 Repairing the Autosampler ...
Страница 198: ...198 Agilent 1200 Series Autosampler Reference Manual 5 Parts and Materials ...
Страница 232: ...232 Agilent 1200 Series Autosampler Reference Manual 7 Introduction to the Autosampler ...
Страница 277: ......