118
Agilent 1200 Series Autosampler Reference Manual
4
Repairing the Autosampler
3 Remove the stator head and stator face.
4 Place the stator face in place on the stator head. Ensure
the pins on the stator engage in the holes in the stator
head.
5 Install stator head and stator face. Tighten the bolts
alternately two turns at a time until the stator head is
secure.
6 Reconnect the pump capillaries to the valve ports (see