Agilent 1200 Series Autosampler Reference Manual
141
Repairing the Autosampler
4
Sampling Unit
Frequency
When accessing internal parts, or when defective.
Tools required
1/4 inch wrench (supplied in accessory kit).
Flat-head screwdriver.
Parts required
Sampling unit G1329-60008 for G1329A
Sampling unit G2260-60008 for G2260A
The sampling units come without injection valve and analytical head
assembly (see
Preparations for this
procedure
Remove the front cover.
Remove the vial tray.
Remove the top cover (
Remove the Illumination assembly (
)
Remove the transport assembly (
1 Remove the pump and column thermostat capillaries
(ports 1 and 6) from the injection valve.
2 Pull out the waste tube from the holder in the leak tray.