138
Agilent 1200 Series Autosampler Reference Manual
4
Repairing the Autosampler
Illumination Assembly
1
Remove the top cover, top plate, and foam (see
2
Disconnect the cable of the illumination assembly at the autosampler
mainboard.
Frequency
Defective LEDs
Defective internal parts (before removing the transport assembly)
Tools required
None
Parts required
Illumination assembly G1367-60040
1 The illumination assembly is located at the top rail of
transport assembly. After disconnecting the cable from
the autosampler mainboard, the illumination assembly
can be removed by turning the rail approximately 60
degrees.
2 Then lift the complete assembly.
To install a new illumination assembly, position it at the
top rail of the transport assembly. There are no screws, the
foam will keep the assembly in place.
Содержание 1200 series
Страница 1: ...A Agilent 1200 Series Standard and Preparative Autosamplers Reference Manual ...
Страница 12: ...12 Agilent 1200 Series Autosampler Reference Manual Contents ...
Страница 42: ...42 Agilent 1200 Series Autosampler Reference Manual 1 Installing the Autosampler ...
Страница 104: ...104 Agilent 1200 Series Autosampler Reference Manual 3 Troubleshooting and Test Functions ...
Страница 108: ...108 Agilent 1200 Series Autosampler Reference Manual 4 Repairing the Autosampler Figure 14 Using the ESD Strap ...
Страница 174: ...174 Agilent 1200 Series Autosampler Reference Manual 4 Repairing the Autosampler ...
Страница 198: ...198 Agilent 1200 Series Autosampler Reference Manual 5 Parts and Materials ...
Страница 232: ...232 Agilent 1200 Series Autosampler Reference Manual 7 Introduction to the Autosampler ...
Страница 277: ......