164
Agilent 1200 Series Autosampler Reference Manual
4
Repairing the Autosampler
autosampler. When the 10-character serial number is entered, press
Enter
to highlight the complete serial number. Then, press
Done
(
F6
).
6
Turn the autosampler off, then on again. The
Records
screen should display
the correct serial number for this module.
7
If a ChemStation is also connected, re-boot the ChemStation now as well.
N O T E
For firmware revisions below A02.00 it is very important never to press Done if the Serial#
box is blank. In this case, the module can no longer be recognized by either the control
module or the ChemStation. The main board must then be replaced.