Agilent 1200 Series Autosampler Reference Manual
161
Repairing the Autosampler
4
Where: XXXXX is the 5-character product number of the module
(e.g. G1314A).
2
Turn OFF the module, then ON again. Then, restart the
Agilent ChemStation. If the serial number you have just entered is different
than the original module serial number, you will be given the opportunity to
edit the configure
1200 access
screen during the restart of the Agilent
ChemStation.
3
After restart, the serial number/type you have just entered can be seen
under the
Instrument
menu of the main user interface screen.
Using the Instant Pilot G4208A
1
Connect the Instant Pilot to the module. Turn ON the module.
2
On the Instant Pilot’s Welcome screen, press
More
, then select
Maintenance
.
Using the
up/down arrows
, select the module where you have to change the
product number or serial number.
3
Press
PN/SN
. This will display a screen where you can enter the product
number and/or serial number.
4
Make your changes, using the information from the product label of your
module.
5
Press
OK
to highlight the complete command.
6
Press
Done
to transfer the information into the main board’s memory. Press
Cancel
quit the process.
7
Turn the module OFF then ON again. The Maintenance screen should
display the correct serial number for this module.
WA R N I N G
If you enter the wrong type, your module will not be accessible anymore. In such
a case see
“Using the Instant Pilot G4208A”
on page 161 or
WA R N I N G
If you enter the wrong type, your module might not be accessible anymore with
the Agilent ChemStation and the Instant Pilot (unsupported module). In such a
case follow the
Содержание 1200 series
Страница 1: ...A Agilent 1200 Series Standard and Preparative Autosamplers Reference Manual ...
Страница 12: ...12 Agilent 1200 Series Autosampler Reference Manual Contents ...
Страница 42: ...42 Agilent 1200 Series Autosampler Reference Manual 1 Installing the Autosampler ...
Страница 104: ...104 Agilent 1200 Series Autosampler Reference Manual 3 Troubleshooting and Test Functions ...
Страница 108: ...108 Agilent 1200 Series Autosampler Reference Manual 4 Repairing the Autosampler Figure 14 Using the ESD Strap ...
Страница 174: ...174 Agilent 1200 Series Autosampler Reference Manual 4 Repairing the Autosampler ...
Страница 198: ...198 Agilent 1200 Series Autosampler Reference Manual 5 Parts and Materials ...
Страница 232: ...232 Agilent 1200 Series Autosampler Reference Manual 7 Introduction to the Autosampler ...
Страница 277: ......