Agilent 1000A Series Oscilloscopes Service Guide
Table 15
Time Scale Accuracy
Limits
Measured time scale error (ppm)
Pass/Fail
±50 ppm from 0 °C to 30 °C
±50 ppm + 2 ppm per °C from 30 °C to 45 °C
+ 5 ppm * (years since manufacture)
Table 16
Trigger Sensitivity
Test Limits
Ch1
Ch2
C3
Ch4
≥
5 mV/div: 1.5 division at maximum bandwidth
≥
5 mV/div: 1 division at 10 MHz
<
5 mV/div: 1.5 division at 20 MHz
<
5 mV/div: 1 division at 10 MHz