Service
1
Agilent 1000A Series Oscilloscopes Service Guide
21
To test analog bandwidth, maximum frequency
C A U T I O N
Ensure that the input voltage to the oscilloscope never exceeds 300 Vrms.
N O T E
This procedure is the only acceptable method for testing the bandwidth of a 1000A Series
oscilloscope.
Table 5
Analog Bandwidth Specification
Analog Bandwidth (-3 dB)
DSO102xA
200 MHz
DSO101xA
100 MHz
DSO100xA
60 MHz