Service
1
Agilent 1000A Series Oscilloscopes Service Guide
15
Calibration
1
Push the
[Utility]
key on the front panel.
2
In the Utility menu, press
Self-Cal
.
3
Follow the on- screen instructions.
N O T E
Let the oscilloscope warm up before testing.
The oscilloscope under test must be warmed up (with the oscilloscope application running)
for at least 30 minutes prior to the start of any performance test.