![Agilent Technologies 1000A Series Скачать руководство пользователя страница 28](http://html.mh-extra.com/html/agilent-technologies/1000a-series/1000a-series_service-manual_2868554028.webp)
1
Service
28
Agilent 1000A Series Oscilloscopes Service Guide
For example, if Pmeas = 3.65 mW:
23
Calculate the gain at the maximum frequency using the expression and
record it in the
“Performance Test Record”
on page 35:
For example, if (Vout @ Max Frequency) = 371 mV, (Vin @ Max
Frequency) = 427 mV and Gain @ 1 MHz = 0.993, then:
Record this value in the Calculated Gain @Max Freq column in the
Analog Bandwidth - Maximum Frequency Check section of the
“Performance Test Record”
on page 35. To pass this test, this value
must be greater than - 3.0 dB.
24
For the remaining channels, move the power splitter to the channel and
repeat steps 4 through 23.
Vin
max
3.65 mW
50
Ω
×
427 mV
=
=
Gain
max
20 log
10
Vout
max
(
)
Vin
max
(
)
⁄
Gain
1
MHz
----------------------------------------------------------
=
Gain
Max
Freq
20 log
10
371 mV 427 mV
⁄
0.993
-------------------------------------------- -1.16
dB
=
=