For each of channel 1’s vertical sensitivities in the DC Gain Test section
of the
“Performance Test Record”
on page 35:
a
For the positive (+) power supply setting:
i
Record the DMM voltage reading as VDMM+.
ii
Record the oscilloscope Vavg reading as .
b
For the negative (- ) power supply setting:
i
Record the DMM voltage reading as VDMM- .
ii
Record the oscilloscope Vavg reading as VScope- .
c
Calculate the DC Gain using the following expression and record this
value in the DC Gain Test section of the Performance Test Record:
14
Repeat
step 15
through
step 23
for channel 2 on two- channel
oscilloscope models or for channels 2, 3, and 4 on four- channel models.
15
Set the power supply voltage to +6 mV.
16
Move the BNC cable from the previously tested channel to the channel
you are currently testing.
Figure 3
Vavg Menu Item
Vavg
measurement
DCGain
=
Δ
V
out
Δ
V
in
--------------- =
V
scope
+
–
V
scope
-
V
DMM
+
–
V
DMM
-
--------------------------------------------