![Agilent Technologies 1000A Series Скачать руководство пользователя страница 22](http://html.mh-extra.com/html/agilent-technologies/1000a-series/1000a-series_service-manual_2868554022.webp)
1
Service
22
Agilent 1000A Series Oscilloscopes Service Guide
Table 6
Equipment Required for Analog Bandwidth Test
Description
Critical Specifications
Recommended
Model/Part Numbers
Signal Generator
100 kHz to 1 GHz at 200 mVrms, 0.01 Hz
frequency resolution, jitter: < 2 ps
Agilent 8648A,
Agilent E4400B, or
Agilent N5181A.
Power Meter
Agilent E-series with power sensor
compatibility
Agilent E4418B
Power Sensor
100 kHz to 1 GHz ±3% accuracy
Agilent 8482A
Power Splitter
Outputs differ by < 0.15 dB
Agilent 11667B
SMA Cable
SMA (m) to SMA (m) 24 inch
50
Ω
BNC Feed
Through Adapter
50
Ω
BNC (f) to BNC (m) feed through
terminator
Agilent 0960-0301
Type-N to SMA
Adapter
Type-N (m) to SMA (f)
Agilent 1250-1250
SMA to BNC Adapter
SMA (m) to BNC (m)
Agilent 1250-0831