Agilent 1000A Series Oscilloscopes Service Guide
27
18
Calculate the reference gain as follows:
Record this value in the Calculated Gain @ 1 MHz column of the
“Performance Test Record”
on page 35.
19
Change the signal generator frequency to the value for the model being
tested as shown in the table below.
20
Change the oscilloscope time base to the value for the model being tests
as shown in the table above.
21
Using the Vpp reading, calculate the Vrms value using the following
expression and record it in the
“Performance Test Record”
on page 35:
For example, if Vpp = 1.24 V:
22
Using the power meter reading, convert this measurement to Volts RMS
using the expression and record it in the
“Performance Test Record”
on
page 35:
Table 7
Oscilloscope Models and Signal Generator Frequency
Setting
Model
DSO102xA
DSO101xA
DSO100xA
Frequency
200 MHz
100 MHz
60 MHz
Time Base
2 ns/div
5 ns/div
10 ns/div
Gain
1
MHz
Vout
1
MHz
Vin
1
MHz
----------------------------
=
Vout
max
Vpp
max
2 2
-------------------
=
Vout
max
1.05
2 2
----------
1.05
2.828
-------------
371 mV
=
=
=
Vin
max
P
meas
50
Ω
×
=
Содержание 1000A Series
Страница 1: ...s1 Agilent 1000A Series Oscilloscopes Service Guide ...
Страница 6: ...6 Agilent 1000A Series Oscilloscopes Service Guide Contents ...
Страница 8: ...8 Agilent 1000A Series Oscilloscopes Service Guide Figures ...
Страница 10: ...10 Agilent 1000A Series Oscilloscopes Service Guide Tables ...
Страница 44: ...44 Agilent 1000A Series Oscilloscopes Service Guide Index ...
Страница 45: ......