Programming the SCSI Device Address
The Small Computer
Systems Interface
(SCSI)
The Small Computer Systems Interface (SCSI) Device Address
designates a specific location or destination of information. During
manufacturing, the SlideWriter's Device Address is preset (default
settings) using DIP switches as part of the SlideWriter's
communication protocol with the host processor.
The SCSI default address is set to 2 (See Table 5-2.)
Verify the SCSI address setting by selecting Function 5 (described
on the following page). The current address will appear in the
Current Exposure Display.
The default settings may be altered in the following way:
• Turn off the SlideWriter. Re-setting the DIP switches does not
affect the SlideWriter until the next power-up, so the unit should be
off when the settings are changed.
• Lift up the left edge of the black plastic strip just beneath the
Display Panel by carefully using a small screwdriver. This strip is the
same width as the Display Panel by about one inch high and
adheres with velcro.
• Pull the left edge of the strip outward far enough to locate the
DIP Switch underneath. When received, it is set to "00010". ("0" is
off, or down, and "1" is on, or up.) Switches 1, 2, and 3 are used
to set the Device Address.
Figure 5-3. Location of the SCSI DIP Switch