Maintenance
Mini Med Series
0000061122
Weekly Cleaning
1) Remove the developer, fixer and wash racks. To prevent chemical contamination, wash
off each rack separately with clean, lint-free cloths and warm water. Clean each roller over
its entire surface. Use isopropyl alcohol if necessary to remove traces of adhesives.
NOTE
: Soft scrub pads, such as nylon net over sponge, work well on rollers. Metallic, or
non-metallic, scrub pads such as "Scotchbrite", must not be used on rollers as they will
damage the roller surface.
2) Inspect each rack thoroughly. Verify that the rollers turn freely and that all guides and
baffles are properly in place. Carefully set each rack aside to drain and dry while you are
cleaning the rest of the processor.
3) Clean the outside surfaces of the processing tank, using warm water with a sponge or
non-metallic scrub pad.
Caution: Never use steel wool on any part of the processor as its residue may cause
rust to form on the metallic parts of the processor.
4) Clean the dryer rollers and rack parts with a damp cloth and wipe dry.
5) Replace all removed racks and other parts.
4-4
Содержание Mini-Medical Series
Страница 4: ......
Страница 10: ......
Страница 12: ......
Страница 14: ......
Страница 16: ......
Страница 26: ...Introduction Mini Med Series 0000061122 ...
Страница 28: ......
Страница 30: ......
Страница 44: ...Installation Mini Med Series 0000061122 Notes 2 14 ...
Страница 46: ......
Страница 56: ......
Страница 66: ......
Страница 98: ......
Страница 120: ...Parts Mini Med Series 0000061122 6 22 Notes ...
Страница 122: ......
Страница 129: ...Accessories Mini Med Series 0000061122 Notes 7 7 ...
Страница 130: ......
Страница 131: ......
Страница 132: ...AFP Imaging Corp 250 Clearbrook Road Elmsford N Y 10523 914 592 6100 Copyright AFP Imaging Corporation 2001 ...