Operation
Mini Med Series
0000061122
Processing Film
Feed film into the processor. As it actuates the feed sensor, the WAIT lamp on the
control panel will light.
If you turn the room lights on after feeding film, to prevent fogging the end of your
film, do not open the daylight cover until after the Wait light goes out or the beeper
sounds.
After the processor completes its processing cycle it will automatically return to Stand-
By mode.
Shutdown and Daily Cleaning
Basic care of the processor goes hand-in-hand with its operation. Following each day’s
work, allow 15 minutes to clean the processor as described below.
Drain Wash Tank
Open the wash tank drain valve and allow the wash water to drain. Rinse out the wash
tank with fresh warm water, then close the drain valve.
Clean Top Cover,
Guides & Rollers
Using a separate wet cloth for developer and fixer, wipe the exposed rollers on each
rack.
Caution: Do not use the same cloth for fixer and developer racks. Fixer may contami-
nate the developer.
Wipe off Processor
Thoroughly wipe the inside and outside surfaces of the top cover and side panels with
a damp cloth. Replace the top cover, leaving a slight opening over the drive shaft to
prevent condensation of chemistry vapors.
3-5
Содержание Mini-Medical Series
Страница 4: ......
Страница 10: ......
Страница 12: ......
Страница 14: ......
Страница 16: ......
Страница 26: ...Introduction Mini Med Series 0000061122 ...
Страница 28: ......
Страница 30: ......
Страница 44: ...Installation Mini Med Series 0000061122 Notes 2 14 ...
Страница 46: ......
Страница 56: ......
Страница 66: ......
Страница 98: ......
Страница 120: ...Parts Mini Med Series 0000061122 6 22 Notes ...
Страница 122: ......
Страница 129: ...Accessories Mini Med Series 0000061122 Notes 7 7 ...
Страница 130: ......
Страница 131: ......
Страница 132: ...AFP Imaging Corp 250 Clearbrook Road Elmsford N Y 10523 914 592 6100 Copyright AFP Imaging Corporation 2001 ...