MAINTENANCE
9.1
Cleaning
The DILUMAT S® does not require strict daily maintenance.
To be able to work in sterile conditions, you must autoclave the distribution tubes in
moist heat.
The bag holder and bowl holder must be removed from the device and cleaned daily
with a damp cloth and hot water. Do not use a scouring pad on the bag holder.
The cover must also be cleaned daily.
Cleaning with 70% alcohol should be done weekly.
In the event of a spillage, be careful not to let liquid run over the sensor, even though
the bag holder has been specially designed to limit the risk of leaks.
9.2
Accessing the Counters
Press the "Validation" key
when powering on the
device.
A E S
L a b o r a t o i r e
D I L U M A T
S
V 1 . 1 2 b
The following screen
appears.
0 0 0
Press the "Validation" key
again to access the counters
for the device.
S / N :
3 1 2 3 4 5 6 7
( 1 ) :
0 0 0 0 0 0 1 2 3
( 2 ) :
0 0 0 0 0 0 1 2 3
( 3 ) :
0 0 0 0 0 0 : 0 0
S/N = serial number
(4)
= Number of dilutions performed
(5)
= Number of dispensings performed
(6)
= Number of hours and minutes powered on
To return to operating mode,
press the "Cancel" key.