DILUMAT S®
MA108201-C
64
10.
Press OK
. And wait until the hourglass disappears.
11.
Check that the
symbol appears, indicating that adjustment is compliant.
12.
Press once more OK
to go back to the menu and print the adjustment report when
a printer is connected to the diluter.
"OK" indicates that the verification is valid; weighing error less than 1%.
The exclamation point indicates that an error occurred during the verification;
unstable weight, error greater than 1%, abnormal gap between the standard weight
and the weighed amount.
Press OK to validate the verification and return to the "Ready to dispense / dilute"
screen.
A verification report is sent to the printer or PC (if one of these options is installed).
Press the Cancel key to return to the "Settings" screen without saving the new data.
A verification report is sent to the printer or PC (if one of these options is installed).