SETUP
5.1
Installation
Recommendations:
The DILUMAT S® must be set up near a power outlet.
The diluter must sit on a flat, stable, and level surface.
It should be used on a clean and dry work surface.
The placement of the device should allow the on/off switch to be
accessible.
5.2
Power
110-120V
220-240V
Voltage
115V +/-10%
230V +/- 10%
Maximum Current
1A 0.5A
Frequency
50-60Hz 50-60Hz
Power Consumed
60 W
60 W
Fuse Rating
1 AT
1 AT
Before plugging in the device, check that the voltage selector on the back is correctly set
for the country’s power characteristics (triangular cursor pointing to the selection).
At delivery, the cursor is set up for 230V +/-10%.
5.3
Turning On
Before turning on the device, it is recommended that the device remain at
room temperature for 30 minutes to stabilize the sensor.
Connect the power cord to the DILUMAT S®.
If using a printer, PC, or bar code
reader, please check that the peripheral
is connected to the Dilumat S® (connection
at the back of the device).
Press the ON/OFF key to turn it on.
A E S
L a b o r a t o i r e
D I L U M A T
S
V 1 . 1 2 b
An initialization sequence displays.