DILUMAT S®
MA108201-C
63
8.5
Adjustment
Note : The adjustment must be done cold. No dilution should be done beforehand.
Use the same reference weight for the adjustment and checking sequences.
1.
Press the "Settings" key.
2.
Using the right/left
arrows, select the
“Adjustment” icon
and press OK
.
1 3 : 1 5
2 0 0 8 / 1 1 / 1 2
0 1
3.
Make sure that the reference weight is actually placed on the diluter.
The
pictogram indicates
that the device is in
“adjustment” mode.
4.
Press the “Tare” key to
start adjustment.
5 0 0 . 0 4 g
0 1
0
g
0 . 0
5.
Place the reference weight on the diluter and check that small characters indicate the
reference mass. Big characters indicate the measured mass.
Example: with a 500g reference mass:
When if you realize an adjustment with a 200g modify the small characters using the arrows
and change the reference weight’s mass. Remember confirming it with the OK key.
By default, the standard weight is 500.00g. It is possible to enter a standard weight from
200.00 to 1500.00g.
The most recent value entered is kept.
6.
Press OK to start the measurement. Wait for the end of the calculation and the
symbol display.
7.
Remove the reference weight from the holder.
8.
Press the "Tare" key . Wait until the hourglass disappears and the symbol
reappears.
9.
Place the reference weight back on he holder.
5
0
0
.
0
4 g
¦ 0 1
0 0
.
0
5
4
Содержание Dilumat S
Страница 1: ...1...
Страница 2: ...DILUMAT S MA108201 C 2...
Страница 9: ...DILUMAT S MA108201 C 9 3 2 Assemblage 1 3 A B 2...
Страница 36: ......
Страница 37: ...DILUMAT S MA108201 C 37...
Страница 38: ...DILUMAT S MA108201 C 38 ENGLISH...
Страница 39: ...DILUMAT S MA108201 C 39...
Страница 46: ...DILUMAT S MA108201 C 46 3 2 Assembly 1 3 A B 2...