Aeroneb
®
Solo System Instruction Manual
19
Aeroneb Solo Aerosol Flow Rate Calculation
Flow rates may vary between individual Aeroneb Solo nebulizers. The minimum flow
rate for all Aeroneb Solo nebulizers is 0.2 mL per minute. In order to calculate the flow
rate of an individual Aeroneb Solo nebulizer:
1. Transfer 0.5 mL of normal saline (0.9%) or intended drug into the Aeroneb Solo
medication cup.
2. Turn on the nebulizer.
3. Using a stop-watch, measure the length of time it takes from the start of
nebulization until all the saline/drug has been nebulized.
4. Calculate the flow rate using the following equations:
Flow rate in mL/min =
Volume of normal saline or drug
Nebulization time in seconds
x 60
)
)
Flow rate in mL/hr =
Volume of normal saline or drug
Nebulization time in seconds
x 60
)
)
x 60
)
)