Aeroneb
®
Solo System Instruction Manual
31
Guidance and manufacturer’s declaration – electromagnetic immunity
This Aeroneb Solo nebulizer system is intended for use in the electromagnetic environment
specified below. The customer or the user of the Aeroneb Solo nebulizer system should assure
that it is used in such an environment.
Immunity Test
IEC/EN 60601
Test Level
Compliance Level
Electromagnetic Environment -
Guidance
Electrostatic
discharge (ESD)
IEC/EN
61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
Transient/burst
IEC/EN
61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
±2 kV for power
supply lines
±1 kV for input/
output lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC/EN
61000-4-5
±1 kV line(s) to
line(s)
±2 kV Line(s) to
earth
±1 kV line(s) to line(s)
±2 kV Line(s) to earth
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5% Ut
(>95% dip in Ut)
for 0.5 cycle
40% Ut(
60% dip in Ut)
for 5 cycles
70% Ut
(30% dip in Ut)
for 25 cycles
<5% Ut
(>95% dip in Ut)
for 5 sec
<5% Ut
(>95% dip in Ut) for
0.5 cycle
40% Ut
60% dip in Ut)
for 5 cycles
70% Ut
(30% dip in Ut)
for 25 cycles
<5% Ut
(>95% dip in Ut)
for 5 sec
Mains power quality should be
that of a typical commercial or
hospital environment. If the user of
the Aeroneb Solo nebulizer system
requires continued operation
during power mains interruption, it
is recommended that the Aeroneb
Solo nebulizer system be powered
from an uninterruptible power
supply or battery.
Power
frequency
(50/60 Hz)
Magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a
typical commercial or hospital
environment.
Note: Ut is the A.C. mains voltage prior to application of the test level.