9
Recharging the Battery
To recharge the battery, connect the AC/DC adapter to the controller and connect
to AC power source. The battery status indicator is amber while charging and green
when fully charged.
If the controller is placed in long-term storage, it is recommended that the battery be
recharged every 3 months.
Allow a minimum of four hours for the internal battery to fully recharge.
Cleaning the Pro-X Controller
Cleaning of controller and controller cable, AC/DC adapter and mounting brackets:
1. Wipe clean with an alcohol based disinfectant wipe or a quaternary ammonium
compound based disinfectant wipe.
2. Check for exposed wiring, damaged connectors, or other defects and replace
controller if any are visible.
3. Visually inspect for damage and replace the controller if any damage is observed.
Warnings
•
Do not immerse or autoclave the Aeroneb Pro-X Controller, cable or AC/DC
adapter.
•
Do not place the Aeroneb Pro-X in an incubator during use.
•
Do not use abrasive or sharp tools.
•
Do not spray liquid directly onto the controller.
•
Do not wrap the nebulizer cable tightly around any of the system components.
•
Do not use in the presence of devices generating high electromagnetic fields
such as magnetic resonance imaging (MRI) equipment.
•
The Aeroneb Pro-X controller contains a nickel metal hydride (NiMH) rechargeable
battery, which should be disposed of in accordance with local governing
regulations at the end of its useful life.
•
Follow local laws and recycling plans regarding disposal or recycling of
components, batteries and packaging.