6
Aerogen
Assembly & Installation
Aeroneb Solo System Set-Up
Perform a functional test of the Aeroneb Solo before use as described in the
Functional Test section of this manual (See page 18).
Figure 2.
Assembly of Aeroneb Solo System
1. Connect the Aeroneb Solo to the T-piece by pushing the nebulizer firmly onto
the T-piece.
2. Insert the Aeroneb Solo and the T-piece into the breathing circuit with the
arrow on the T-piece pointing in the direction of the air-flow within the circuit.
Note:
For use with other accessories, refer to Figure 8, Figure 9 and Figure 10.
3. Connect the Aeroneb Pro-X controller to the Aeroneb Solo using the nebulizer
cable.
4. To operate on AC power (the primary mode of operation), connect the Aeroneb
Pro-X AC/DC adapter to the Aeroneb Pro-X controller.
5. Plug the adapter into an AC power source.
6. Aeroneb Pro-X can be battery-operated for portable applications. The
rechargeable battery can power the System for up to 45 minutes when fully
charged. In the case of AC power failure the controller will automatically switch
to battery operation.