74
4.2.18 Measurement of Reactive AC+DC Power
Measurement
Range (2) (4)
5 to
9999 var
10.00 to
99.99 kvar
100.0 to
999.9 kvar
1000 to
1500 kvar
(1)
Specified
Measurement
Range
1 to 100% of the
measurement range
0 to 100% of the measurement range
Accuracy (3)
(7)
1000A
± (2% R +10cts)
1000 to 2000A
± (2.5% R +10cts)
1000A
± (2% R +3cts)
1000 to 2000A
± (2.5% R +3cts)
Resolution
1 var
10 var
100 var
1 kvar
Note (1)
- Display of OL above 1500 kvar in single-phase (1000V x 1500A).
- Bandwidth in AC in voltage = 3 kHz, in current = 2 kHz
Notes (2), (3) , (4), (5), (6) and (7)
of the previous § apply
Specific characteristics in MAX/MIN mode in power
(from 10Hz to 1kHz in AC and AC+DC):
Accuracy: add 1% R to the values in the table above.
Capture time: approximately 100ms
4.2.19 Calculation of the Power Factor (PF)
Measurement Range (1)
0.00 to +1.00
Specified Measurement
Range
0 to 50% of the
measurement range
50 to 100% of the
measurement range
Accuracy (2)
± (3% R +3cts)
± (2% R +3cts)
Resolution
0.01
Note (1)
If one of the terms in the calculation of the power factor is displayed
as "OL", or forced to zero, the display of the power factor is an
indeterminate value "----".
Note (2)
In balanced three-phases, with deformed signals (THD and
harmonics), uncertainties are guaranteed since Ф > 30°. Additional
errors are following, depending of THD:
Add +1% for 10% < THD < 20%
Add +3% for 20% < THD < 30%
Add +5% for 30% < THD < 40%
NOTE:
The PF is always positive
Specific characteristics in MAX/MIN mode (from 10Hz to 1kHz):
Accuracy: add 1% R to the values in the table above.
Capture time: approximately 100ms.