32
3.10
POWER MEASUREMENTS W, VA, VAR AND PF
This measurement is possible in single-phase or in balanced 3-phase.
NOTE:
If performing DC or AC+DC power measurements, correct the DC zero in
current first (see § 3.8.2)
For the power factor (PF) and VA and var, the measurements possible are only
available in AC or AC+DC modes.
3.10.1 Single-Phase Power Measurement
1.
Set the switch to
and select
VA
,
var
, or
PF
by pressing the
button until the desired choice is reached.
2.
The device automatically displays
AC+DC
. To select AC, DC, or
AC+DC, press the
(yellow button) until the desired choice is
reached.
3.
Connect the black lead to the
COM
terminal and the red lead to the "
+
"
terminal.
4.
Connect the test probes or the alligator clips of the black lead on the
neutral (N), then those of the red lead on the L phase.
5.
Clamp around only the corresponding conductor, respecting the
direction.
The measurement is displayed on screen.
Display
Quantity
1st row
Active power W (DC, AC or AC+DC)
2nd row
Reactive power var (AC or AC+DC)
3rd row
Apparent power VA (AC or AC+DC)