30
In AC and AC+DC:
Display
Quantity
1st row
Total RMS current in A RMS or TRMS
2nd row
Crest factor (CF)
3rd row
DC current component A
DC
3.9
STARTING CURRENT OR OVERCURRENT (True InRush
™
)
MEASUREMENT
NOTE:
The measurement can be made only in AC or DC mode (AC+DC mode
disabled).
To measure a starting current or overcurrent, proceed as follows:
1.
Set the switch to
, correct the DC zero (see §3.9.2), then clamp the
jaws around the conductor to be measured.
2.
Perform a long press on the
button. The
InRh
symbol is
displayed, along with the triggering threshold. The clamp then awaits
detection of the True InRush
™
current. "------" is displayed and the A
symbol flashes.
3.
After detection and acquisition for 100 ms, the RMS value of the True
InRush
™
current is displayed. Pressing the
button will display the
PEAK+/PEAK- values subsequently.
4.
A long press on the
button or a change of function on the rotary
switch will exit the True InRush
™
mode.
NOTE:
The triggering threshold in A is 20A if the initial current is zero (starting of
installation). For an established current (overload in an installation) see §3.4.2.
Display
Quantity
1st row
“Inrh”
2nd row
True Inrush value in A
3rd row
Triggering threshold in A