69
4.2.8
Calculation of the Crest Factor (CF)
Measurement range
1.00 to 3.50
3.51 to 5.99
6.00 to 10.00
Specified
Measurement Range
(from 5V or 5A)
0 to 100% of the measurement range
Accuracy (zero
corrected in A
DC
)
± (2% R +2cts)
± (5% R +2cts)
± (10% R +2cts)
Resolution
1ct
NOTE:
Peak values limited to 1500V or 1500A
4.2.9
Calculation of the RIPPLE in DC
Measurement range
0.1 to 99.9%
100.0 to 1000%
Specified Measurement
Range (from 3A
DC
and
2V
DC
)
2 to 100% of the
measurement range
0 to 100% of the
measurement range
Accuracy
± (5% R +10cts)
Resolution
0.1
NOTE:
If one of the terms for the calculation of the RIPPLE is displayed as "OL",
or forced to zero, the RIPPLE displayed is an indeterminate value, "----".
4.2.10 Continuity Measurement
Measurement Range
0.0 to 999.9
Open-circuit Voltage
3.6V
Measurement Current
550µA
Accuracy
± (1% R +3cts)
Buzzer Triggering Threshold
40
4.2.11 Resistance Measurement
Measurement Range (1)
0.0 to
999.9
1000 to
9999
10.00 to
99.99k
Specified Measurement
Range
1 to 100% of the
measurement range
0 to 100% of the
measurement range
Accuracy
± (1% R +3cts)
Resolution
0.1
1
10
Open-circuit Voltage
3.6V
Measurement Current
550µA
100µA
10µA
Note (1)
Above the maximum display value, the display unit indicates "OL".
The "-" and "+" signs are not managed.