74
Power Quality Analyzer PowerPad
®
IV Model 8345
programmed time, if the SD card has enough space
Press the
button in the function bar to start the capture of a current
programmed in the configuration (§ 3.4.6) at the end of the current minute
plus one minute
Press the button in the function bar to suspend the in-progress recording
Press the button in the function bar to change the voltage, current, or
shock wave thresholds
The capture will begin recording when the current exceeds the threshold.
11.2 THE LIST OF CAPTURES
Press
to view the captures performed
Each entry in the recording list includes the recording’s name, start date,
start time, end date, and end time
Press to erase the selected capture
To erase all of the inrush current captures, refer to § 3.3.4.
If the end date is red, the recording could not continue to the planned end date
because of:
■
A power supply problem (the instrument turned off due to low battery)
■
A write error on the SD card
To learn what the error number means, use the help button
.
11.3 READING A CAPTURE
Select the capture from the list, and press the confirm
button to open it
Captures with red end dates may be unusable.
Each entry in the capture list includes the capture’s name, number of inrush
To view different pages