2.2.12 Measuring Dual Parameters
2-59
15. Press FREQ, 1, 8, 0, 0 and MHz.
A center frequency of 1800 MHz is set for the lower screen.
16. Press COUPLE, VBW AUTO/MNL, 1, 0, 0 and kHz.
A VBW of 100 kHz is set for the lower screen.
17. Press SWP and Trigger Setup.
The Trigger Setup dialog box is displayed.
18. Set Source to VIDEO.
The trigger source is set to VIDEO. The cursor moves to Slope. The trigger level
mark "
→
" is displayed on the left-hand side of the scale.
19. Press Hz(ENTER).
The trigger slope is set to "+" and the cursor moves to Trigger Level.
20. Adjust the trigger level.
Adjust the trigger level to the middle of the burst signal turning the data knob. A
stably triggered display is obtained.
21. Press RETURN.
The Trigger Setup dialog box is removed.
22. Press Sweep Time AUTO/MNL(MNL), 5 and kHz(ms).
A sweep time of 5 msec is set for the lower screen.
Figure 2-70 Displaying the Traces of 900 MHz and 1800 MHz in the Separate Screen Mode
23. Press SRCH.
The level on the lower screen can be measured using the marker.