R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
2.2.13 Calibration
2-64
2.2.13.3 Cal Each Item
Performs one Cal Each Item though Cal All performs all calibration item.
This section describes the PBW calibration.
Setup
Connect the calibration signal.
1.
Connect the N-BNC adapter to the INPUT connector on the front panel.
2.
Connect the Input cable between CAL OUT and INPUT connectors on the front
panel.
Performing the calibration
3.
Press SHIFT and 7(CAL).
The Cal menu used for calibration appears (See Figure 2-72).
4.
Press Cal Each Item and PBW.
PBW (noise power bandwidth) calibration is performed.