R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
6.3 R3273 Specifications
6-14
6.3
R3273 Specifications
(1)
Frequency
Characteristics
Specification
Frequency range:
Frequency reading accuracy:
±
(Frequency reading
×
Frequency reference ac Span
×
Span accu-
racy + 0.15
×
Resolution ba 10 Hz)
Marker frequency counter
(SPAN < 1 GHz)
Accuracy (S/N > 25 dB):
Delta counter:
Resolution:
±
(Marker frequency
×
Frequency reference ac 5 Hz
×
N + 1LSD)
±
(
∆
Frequency
×
Frequency reference ac 10 Hz
×
N + 2LSD)
1 Hz to 1 kHz
Reference frequency source
stability
Aging:
Temperature stability:
A warm-up (Nominal):
OPT21
Aging:
Temperature stability:
A warm-up (Nominal):
OPT22
Aging:
Temperature stability:
A warm-up (Nominal):
OPT23
Frequency accuracy:
Aging:
Temperature stability:
Warm-up:
±
3
×
10
-8
/day
±
1
×
10
-7
/year
±
1
×
10
-7
Temperature range: 0 to 40
°
C in reference to the frequency
measured at 25
°
C
±
2
°
C
±
5
×
10
-8
/3 min (In reference to the frequency measured 60 min. after the
power-on)
±
5
×
10
-9
/day
±
8
×
10
-8
/year
±
5
×
10
-8
Temperature range: 0 to 40
°
C in reference to the frequency
measured at 25
°
C
±
2
°
C
±
5
×
10
-8
/3 min (In reference to the frequency measured 60 min. after the
power-on)
±
3
×
10
-10
/day
±
2
×
10
-8
/year
±
5
×
10
-9
Temperature range: 0 to 50
°
C in reference to the frequency
measured at 25
°
C
±
1
×
10
-8
/30 min
In reference to the frequency measured 24 hours after
±
5
×
10
-9
/60 min
the power-on at an ambient temperature of 25
°
C
±
5
×
10
-9
±
1
×
10
-10
/month
±
1
×
10
-9
Temperature range:0 to 40
°
C in reference to the frequency
measured at 25
°
C
±
1
×
10
-9
/15 min
100 Hz to 26.5 GHz
18 GHz to 60 GHz (external mixer used, synchonizable with up to 325 GHz)
Frequency
Frequency band
Harmonic order N
100 Hz to 3.5 GHz
0
1
3.5 GHz to 7.5 GHz
1
1
7.4 GHz to 15.4 GHz
2
2
15.2 GHz to 26.5 GHz
3
4
Built-in YIG tuning pre-selector at 3.5 GHz to 26.5 GHz
{