2.3.1 Measuring Average Power of Digital Modulation Signal
2-73
Figure 2-81 Checking the Input Signal
12. Press SPAN and Zero Span.
The frequency span is set to zero (See Figure 2-82).
Figure 2-82 Setting Measurement Conditions for Average Power Measurement
Measuring average power
13. Press POWER and Average Power.
Measures the power averaged over the object range and displays the result.
Allows you to set the averaging count.
14. Press 1, 0 and Hz(ENTR).
An averaging count of 10 is set.
The average power and averaging count set are displayed in the result area (See
Figure 2-83).